Biosystems Bts 350 User Manual -

The BTS 350 user manual and other support documents can be found through various online resources.

: The analyzer supports several measurement methods, including: Fixed Time Differential Mode Quality Control (QC)

: To safely shut down, first clean the aspiration system with distilled water, then navigate to the menu's "shutdown" button before physically turning off the machine. LED Advantage biosystems bts 350 user manual

The system may require incubation time, which is managed by the software.

Internal Peltier element factory-calibrated to 25°C, 30°C, and 37°C ( ±plus or minus Data and Interfaces The BTS 350 user manual and other support

is designed for low maintenance, but regular care ensures longevity.

Direct the opposite end of the tubing into an authorized biohazard waste container. Thermostatic System: Peltier system set at | Display

technical documentation, below are the key technical parameters: Wavelengths: (plus 2 free positions for optional filters). Thermostatic System: Peltier system set at

| Display Code | Meaning | Manual’s Solution | | :--- | :--- | :--- | | | Cuvette missing or dirty | Check cuvette placement. Clean with HCl 1N. Replace if scratched. | | Err 4 | Lamp low intensity | Run lamp diagnostic. If voltage is low, replace lamp. | | Err 7 | Temperature below 36.5°C | Check ambient temp (must be <30°C). Wait 20 min for stabilization. | | Err 10 | Overflow detected | Immediate: Stop pump. Empty waste bottle. Clean overflow sensor. | | No display | Power failure or fuse blown | Unplug unit. Replace glass fuse in power input module. |